User Manual 3155
Programming Reference 4-54
TRIGger:BURSt{OFF|ON|0|1}
Purpose
This command places the 3155 in burst mode. This command affects the 3155 only after it is set to triggered
operation with the command INIT:CONT OFF.
ON or 1 enables burst mode
OFF or 0 disables burst mode
Parameter type
Discrete
TRIGger:BURSt?
Response and default
The 3155 returns 1 if burst mode is on, or 0 if burst mode is off. Default value is 0.
TRIGger:COUNt<count>
Purpose
This command sets the burst counter.
<count> sets the number of burst counts
Parameter type
Numeric (integer only)
Parameter range
<count> 1 to 1 Meg
TRIGger:COUNt?
Response and default
The 3155 returns the present count value. Default value is 1.
TRIGger:GATE{OFF|ON|0|1}
Purpose
This command places the 3155 in gate mode. This command affects the 3155 only after it is set to triggered
mode with the command INIT:CONT OFF.
ON or 1 enables gate mode
OFF or 0 disables gate mode
Parameter type
Discrete
TRIGger:GATE?
Response and default
The 3155 returns 1 if the gate is on, or 0 if the gate is off. Default value is 0.