3155 User Manual
Performance Verification 5-4
2. Set the DMM to ACV measurements (RMS).
3. Connect the Model 3155 output to DMM input. Termi-
nate the output with a 50O feedthrough termination. Set
the amplitude and verify DMM reading as follows:
3155 SETTING DMM READING
7.50 V 2.682 V – 2.622 V
750.0 mV 268.2 mV – 262.2 mV
DC Offset Charac-
teristics
Accuracy Specifications:
± (1% of reading +1% of amplitude +200µV), ± 400mV window
± (1% of reading +1% of amplitude +2mV), ±4V window
Equipment: DMM, 50O feedthrough termination
1. Configure the Model 3155 as follows:
Function: Sine
Frequency: 1.000 MHz
Amplitude: 1 V
Remote Commands: :RES
:OUTP ON
:FUNC SIN
:FREQ 1e6
:VOLT 1
:VOLT:OFFS 2.5
2. Set the DMM to DCV measurements
3. Connect the Model 3155 output to the DMM input. Ter-
minate the output with a 50O feedthrough termination.
Set the Model 3155 offset and verify the DMM reading
as follows:
OFFS SETTING DMM READING
±2.500 V ±2.537 V to ±2.463 V
4. Change the Model 3155 AMPL setting to 100 mV. Set
VOLT:OFFS and verify the DMM reading as follows:
OFFS SETTING DMM READING
±250.0 mV ±253.7 mV to ± 246.3 mV