3155 User Manual
Performance Verification 5-1
Chapter 5
PERFORMANCE VERIFICATION
What’s in This
Chapter
This chapter provides performance tests for the Model 3155
Waveform Synthesizer.
CAUTION
ALWAYS PERFORM DISASSEMBLY, REPAIR AND
CLEANING AT A STATIC SAFE WORKSTATION.
Performance
Checks
The following performance checks verify proper operation of
the instrument and should normally be used:
1. As a part of the incoming inspection of the instrument
specifications;
2. As part of the troubleshooting procedure;
3. After any repair or adjustment before returning the in-
strument to regular service.
Environmental
Conditions
Tests should be performed under laboratory conditions having
an ambient temperature of 25
o
C
±
5
o
C and at relative humidity
of less than 80%. If the instrument has been subjected to con-
ditions outside these ranges, allow at least one additional hour
for the instrument to stabilize before beginning the adjustment
procedure. Always perform a self-test sequence before com-
mencing with the performance checks. The self-test, if exe-
cuted without any failure, ensures proper operation of the syn-
thesizer. If self-test failures have been encountered, the in-
strument first needs to be serviced and the source of failure
removed.
CAUTION
SENSITIVE ELECTRONIC DEVICES
DO NOT SHIP OR STORE NEAR
STRONG ELECTROSTATIC,
ELECTROMAGNETIC, MAGNETIC OR
RADIOACTIVE FIELDS
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Seitenansicht 185
1 2 ... 181 182 183 184 185 186 187 188 189 190 191 ... 203 204

Kommentare zu diesen Handbüchern

Keine Kommentare